Section 4: Basic Configuration/Start-Up
January 2015
Basic Configuration/Start-Up
4.6.5 Using the Setup function
To install a Rosemount 3300 Series Transmitter by using the Setup function, do the following:
1. Start the RCT software.
2. In the RCT workspace select the Setup icon (make sure the Tools area is open),
or
select the View > Setup menu option.
3. Select the appropriate tab:
Info: information about the device.
Basics: information about probe type and measurement units.
Output: information about transmitter variables.
Probe: information about probe type and length.
Geometry: information on reference gauge height, probe length, mounting type,
nozzle inner diameter, and nozzle height.
Environment: information about measurement mode, the upper product dielectric
constant and advanced environment options
Volume: specification of tank geometry for volume calculations.
Display: display panel settings.
Signal Quality Metrics: information about signal quality and surface/noise margin